PVC PANEL FOR MONITOR
SPECIFIC SENSOR Type: Oxytrace G sensor Application: Sensor for measuring dissolved oxygen in ultrapure water Swansensor Oxytrace G precise oxygen measuring cell with integrated temperature sensor and protective electrode for faster initial response after maintenance. Easy maintenance thanks to the sensor cover with integrated membrane. Easy calibration. No exposed metal surfaces, no corrosion in corrosive media. High electrical reliability. No integrated electronics. Technical data: Clark oxygen electrode, gold cathode, silver anode, silver guard, current-free electrode system, robust 25 µm fluoropolymer diaphragm, measuring range: 0-20 ppm O2 (25 °C) or 0-200% saturation, accuracy: 0.3% if calibration temperature equals measuring temperature, 1.5% with a deviation of ±10 °C from calibration temperature, accuracy: ±1% of reading or ±0.15 ppb, response time: t90 < 30 seconds (rising concentration), minimum flow rate: 50 cm/s, pressure resistance: 3 bar, operating temperature: max. 50 °