PVC
Performance standard Cooling principle Peltier cooling Number of gas paths 1 Number of Peltier elements 1 Integrated condensate pump option Integrated condensate monitoring option Operation Gas flow per gas path* max. 250 Nl/hr Gas inlet temperature* max. 140 °C; SS heat exchanger: max. 180 °C Gas inlet dew point* max. 80°C Gas outlet dew point 5 °C (factory default); adjustable from 0,5 °C to 7,5 °C Dew point stability (for constant inlet conditions) ±0,1 K Ambient temperature 5° to 45 °C Cooling capacity total max. 15 W Operating pressure with condensate pump 0,2 to 2,2 bara Max. operating pressure without condensate pump 4,0 bara; SS heat exchanger: max. 19 bara Ready for operation < 30 min Pressure drop at max. flow rate 3 mbar Construction Dimensions over all (W x H x D) 244 x 308 x 211 mm Installation wall mounting Mounting position horizontal Weight** approx. 6,5 kg Housing / Colour stainless steel / natural Gas wetted materials (depending on configuration) PVDF, SS316Ti, FFKM
Price, availability, and delivery times will be confirmed in your quote.
| Manufacturer | JCT |
|---|---|
| Part number | JCM-300 |